MKS 628B Series Capacitance Manometers
![]() |
MKS has improved on the industry standard Type 628A by incorporating new features into the Type 628B Baratron® capacitance manometer.
Features
- Temperature regulated to 100° C
- Accurate to 0.25% of reading
- Includes updated temperature control electronics to provide superior long-term stability and repeatability
- Optional heater and temperature status LED/switches indicate that the heater (which maintains the sensor temperature to 100° C) is at temperature
- Available from 0.05–1,000 mm Hg (Torr)
- Input power is ±15 VDC
- Output signal is 0–10 VDC
- Replaces the Type 628A and Type 128 absolute capacitance manometer
- Inconel® construction for excellent resistance to corrosive gases
Specifications Table
| Measurement Method | Capacitance Manometer |
| Accuracy (% of measurement) | 0.25% (1, 2, 10, 20, 100, & 1,000 Torr FS) 0.5% (0.05, 0.1 Torr FS) |
| Temperature (operating) | 15-50° C |
| Max. Pressure | 45 psia |
| Materials Exposed to Vacuum | Inconel |
| Analog Outputs | 0-10 VDC |
| Internal Volume | 6.3 cc |
| Input Power | ±15 VDC ±5% @ 0.6A max. |
| Temperature Coeff. (zero) | 0.002%(1-1,000 Torr FS) 0.01% (0.1 Torr) 0.02% (0.05 Torr) |
| Temperature Coeff. (span) | 0.02% of reading/° C |
| Full-Scale Pressure | 0.05, 0.1, 1, 2, 10, 20, 100, & 1,000 Torr |
| Resolution | 0.001% of F.S. (0.1 Torr or greater) 0.002% of F.S. ( 0.05 Torr) |