KJLC® Convection Enhanced Pirani Gauges


These gauge tubes and controllers are ideal for new applications or as low-cost drop-in replacements to the Convectron® brand of Granville-Phillips® gauges.

Features

  • New applications or low-cost drop-in replacements for Convectron gauges, Mini-Convectron® modules, and controllers
  • Less prone to damage from mechanical shock or vibration than competitor's tubes
  • Less outgassing and faster response to pressure changes than competitor's tubes
  • Wide measuring range from 1 x 10-4 to 1,000 Torr(1.3 x 10-4 to 1,333 mbar)
  • Controllers feature universal power and are CE marked

KJLC Convection Enhanced Pirani Gauge Comparison Guide

275 Series Gauge Tubes

275 Series Gauge Tubes

300 Series Gauge

375 Series Gauge

275i Series Gauge

275i Series Gauge

375 Series Gauge

300 Series Gauge

Design Gauge Tube and Panel Mount / Benchtop Gauge Controller Gauge Tube and Panel Mount / Benchtop Gauge Controller Gauge with Integrated Controller & Display Gauge with Integrated Controller & Display
Measurement Range 1 x 10-4 to 1,000 Torr (1.3 x 10-4 to 1,333 mbar) 1 x 10-4 to 1,000 Torr (1.3 x 10-4 to 1,333 mbar) 1 x 10-4 to 1,000 Torr (1.3 x 10-4 to 1,333 mbar) 1 x 10-4 to 1,000 Torr (1.3 x 10-4 to 1,333 mbar)
Display See 375 or MGC Controllers 4-digit OLED (Torr, mbar, or Pa) 3-digit LED (Torr & mTorr or Bar & mBar) 4-digit OLED (Torr, mbar, or Pa)
Temperature (Operating) 0-50°C 0-50°C 0-40°C 0-40°C
Accuracy (Typical) 1.0 x 10-4 to 1.0 x 10-3 Torr; 0.1 mTorr resolution
1.0 x 10-3 to 400 Torr; ± 10% of reading
400 to 1,000 Torr; ± 2.5% of reading
1.0 x 10-4 to 1.0 x 10-3 Torr; 0.1 mTorr resolution
1.0 x 10-3 to 400 Torr; ± 10% of reading
400 to 1,000 Torr; ± 2.5% of reading
1.0 x 10-4 to 1.0 x 10-3 Torr; 0.1 mTorr resolution
1.0 x 10-3 to 400 Torr; ± 10% of reading
400 to 1,000 Torr; ± 2.5% of reading
1.0 x 10-4 to 1.0 x 10-3 Torr; 0.1 mTorr resolution
1.0 x 10-3 to 400 Torr; ± 10% of reading
400 to 1,000 Torr; ± 2.5% of reading
Repeatability (Typical) ± 2% of reading ± 2% of reading ± 2% of reading ± 2% of reading
Set-Point Relays Two SPDT Relays Two SPDT Relays One SPDT Relay Two SPDT Relays
Analog Outputs 1) Non-linear S-curve 0.375-5.659 VDC
2) Log-linear 1-8 VDC, 1V/decade
3) Linear 0-10 VDC, user scalable
1) Non-linear S-curve 0.375-5.659 VDC
2) Log-linear 1-8 VDC, 1V/decade
3) Linear 0-10 VDC, user scalable
1) Non-linear analog S-curve 0.375 to 5.659 VDC
2) Log-linear 1 to 8 VDC, 1 V/decade
1) Non-linear S-curve 0.375-5.659 VDC
2) Log-linear 1-8 VDC, 1V/decade
3) Linear 0-10 VDC, user scalable
Input Power See 375 or MGC Controllers 12-28 VDC, 2 W, protected against power reversal and transient over-voltages 12-28 VDC, 2 W, protected against power reversal and transient over-voltages 12-28 VDC, 2 W, protected against power reversal and transient over-voltages
Connections See 375 or MGC Controllers Gauge tube: 9-pin D female

Analog output and digital interface: 9-pin sub-D (male)

Relay outputs: 6-pin pluggable terminal block, 2-pin pluggable terminal block (DC), or optional universal AC power supply
9-pin sub-D (male) 9-pin sub-D (male) and 15-pin high-density sub-D (male)
Learn More Learn More Learn More Learn More