Close

Please select your country or region to be
directed to the appropriate Lesker site.

Material Deposition Chart




+ Show Key of Symbols

A

B

C

D

E

G

H

I

K

N

O

P

R

S

T

U

V

Y

Z

Z-Factors

Empirical Determination of Z-Factor

Unfortunately, Z Factor and Shear Modulus are not readily available for many materials. In this case, the Z-Factor can also be determined empirically using the following method:

  • Deposit material until Crystal Life is near 50%, or near the end of life, whichever is sooner.
  • Place a new substrate adjacent to the used quartz sensor.
  • Set QCM Density to the calibrated value; Tooling to 100%
  • Zero thickness
  • Deposit approximately 1000 to 5000 A of material on the substrate.
  • Use a profilometer or interferometer to measure the actual substrate film thickness.
  • Adjust the Z Factor of the instrument until the correct thickness reading is shown.

Another alternative is to change crystals frequently and ignore the error. The graph below shows the % Error in Rate/Thickness from using the wrong Z Factor. For a crystal with 90% life, the error is negligible for even large errors in the programmed versus actual Z Factor.